JaLCDOI 10.18926/11755
タイトル(別表記) New Stepwedge for Bootstrap Sensitometry in Medical Radiography
フルテキストURL 001_019_027.pdf
著者 吉田 彰| 小橋 利美| 大川 義弘| 三上 泰隆| 平木 祥夫| 東 義晴| 山田 俊治| 杉田 勝彦|
抄録 A new stepwedge has been developed for bootstrap sensitometry, which is used for determining the characteristic curve of radiographic screen/film systems. Since each step of the stepwedge is separated by the lead septum, flare generated under one step will not penetrate the region of another step. The metal of the stepwedge can be changed with various materials. Characteristic curves obtained with new stepwedge bootstrap (NSB) and conventional stepwedge bootstrap (CSB) techniques were compared with those obtained for screen/film systems with inverse sqare sensitometry. The inverse square method was used as the reference standard for accuracy. The NSB method provided characteristic curves that agreed well with those obtained using the inverse square method. The slope of the curve in the CSB method was lower than that in the inverse square or NBS method, because of scatter radiation. The shape of the characteristic curve from the NSB method did not vary with the materials of step metals (aluminum and copper). The cumulative error involved when using the bootstrap method is also discussed.
キーワード Stepwedge Bootstrap Characteristic curves Sensitometry Radiographic screen/film systems
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1991-03-25
1巻
開始ページ 19
終了ページ 27
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313394
JaLCDOI 10.18926/11808
タイトル(別表記) RESOLUTION PROPERTIES OF CRT MONITORS
フルテキストURL 002_081_088.pdf
著者 竹内 知行| 吉田 彰| 東 義晴| 杉田 勝彦| 中村 伸枝| 岸 幹二| 平木 祥夫|
抄録 We have developed a new technique for measuring the modulation transfer function (MTF) of a cathode ray tube (CRT) monitor. The "nominal" MTF was measured by displaying a line image of a width of one pixel on the CRT monitor using the pattern generator. We also measured the effective pixel width. In order to obtain the "actual" MTF, the nominal MTF was divided by the sinc function which is the Fourier transform of the effective pixel width. The results indicate that the corrected MTF converges to the constant value as the effective pixel width is narrower. Thus, the actual MTFs were determined and compared in the horizontal direction for two CRT monitors.
キーワード CRT Monitor Resolution Effective Pixel Width Modulation Transfer Function (MTF)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1992-02-05
2巻
開始ページ 81
終了ページ 88
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313719
JaLCDOI 10.18926/11805
タイトル(別表記) Dose Distributions at Standard Diagnostic X-Ray Energy
フルテキストURL 002_063_071.pdf
著者 三宅 英昭| 稲村 圭司| 中桐 義忠| 門久 繁文| 三上 泰隆| 東 義晴| 山田 俊治| 杉田 勝彦| 平木 祥夫|
抄録 Exposure dose has been indicated by surface dose or transit dose, but they could not indicate dose distributions inside the body. Modulex as the radiation therapy planning system was used for dose distributions at standard diagnostic X-ray energy. X-ray is low energy X-ray at standard diagnostic radiography, so alterations of the energy and the scattering dose distribution by absorbers are quite different from those at high energy X-rays. Mix-DP was put to the homogeneous tissue use, then Bone equivalent phantom or Lung equivalent phanton was put to the inhomogeneous tissue use. Density correction factors for inhomogeneous phantoms were gotten by the calculation of the water equivalent thickness. In Bone the inhomogeneous correction depended strongly on the bone thickness. In Lung it was in need of one density correction factor and the scattering correction method for the decrease of the back scattering. The calculated dose distributions by Modulex agreed with measured data when each correction was carried out, and it was indicated that those data apply to inclinical situations.
キーワード Exposure Dose Dose Distribution Radiation Therapy Planning System Density Correction Scattering Correction
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1992-02-05
2巻
開始ページ 63
終了ページ 71
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313328
JaLCDOI 10.18926/11820
タイトル(別表記) Photon background caused by the reduction of the electron beam energy - Materials of scattering foil -
フルテキストURL 003_013_018.pdf
著者 中桐 義忠| 三村 誠一| 稲村 圭司| 田原 誠司| 三宅 英昭| 東 義晴| 江草 具視| 三上 泰隆| 平木 祥夫| 山田 俊治| 杉田 勝彦|
抄録 The total skin electron beam therapy has been one of the clinical treatment for peripherally T-cell lymphoma; Mycosis fungoides, adult T-cell lymphoma, and so on. The crucial points in this treatment are not only having an optimum energy level of electron beam for a target volume (a tissue) but also keeping the photon back ground low. It is not easy to regulate those points by the control panel, however, for the equipment that is conventinally used for electron beam, theoretically, is to exchange lead (Pb), which is ordinarily used, to a low atomic number material as a scattering foil. We examined several different kinds and / or various thickness as a scattering foil material that can make the electron beam lower without an increase of the contaminant as X-ray. We hereby reported the results, and strongly suggested the following two materials in use; acrylic plate, carbon board, and so on, which are easily available and worked, would be practically useful for the total skin electron beam therapy.
キーワード 電子線エネルギー X線混入率 電子線全身照射法 スキャタリングホイル
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1993-01-31
3巻
開始ページ 13
終了ページ 18
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313375
JaLCDOI 10.18926/11878
タイトル(別表記) Dosimetry in Radiography
フルテキストURL 004_069_073.pdf
著者 中桐 義忠| 後藤 佐知子| 三上 泰隆| 東 義晴| 渋谷 光一| 稲村 圭司| 田原 誠司| 宇野 弘文| 高田 義明| 丹谷 延義| 山田 俊治| 杉田 勝彦| 平木 祥夫|
抄録 Utilization of radiation in medical treatment is increasing more and more ; consequently, It becomes more important to estimate exposure dose correctly. Altough there are many reports about exposue dose for patients, most of them merely describe the results of the measurements by parts of the body. Exposure dose differs with equipment, instruments, screen-firm system, contition of radiography, and so on. This paper describes the relation between skin dose and contitions of radiography, and also shows the result of measurement of "TPR" which needs to know the absorbed dose of each organ.
キーワード 線量測定 低エネルギー 被曝
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1994-01-31
4巻
開始ページ 69
終了ページ 73
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313669
JaLCDOI 10.18926/11886
タイトル(別表記) Analysis of Tempormandibular Joint by Image Processing -Discrimination of Condylar Position-
フルテキストURL 004_075_082.pdf
著者 東 義晴| 後藤 佐知子| 中桐 義忠| 中村 伸枝| 渋谷 光一| 山田 俊治| 杉田 勝彦| 三上 泰隆| 平木 祥夫|
抄録 Tomography is clinically useful in the diagnosis of the temporomandibular joint (TMJ) syndrome. The TMJ is formed by the condyle and the glenoid fossa, and there exists TMJ space between them. Image processing was employed in the present study, and the reference points of the condyle and the glenoid fossa were determined from the tomogram. Subsequently, the TMJ space was measured automatically in the image analysis. The TMJs of 98 patients were examined practically and the characteristics of each result were studied. There is a possibility of discriminating an abnormal condylar position from a normal one by making use of the TMJ space and the distance between two reference points.
キーワード Condylar Position Temporomandibular Joint Space Image Processing Tomogram Discrimination
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1994-01-31
4巻
開始ページ 75
終了ページ 82
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313405
JaLCDOI 10.18926/11874
タイトル(別表記) Estimation of balium preparations with high density and low viscosity
フルテキストURL 004_057_060.pdf
著者 渋谷 光一| 小橋 高郎| 小笠原 光男| 丹谷 延義| 中桐 義忠| 東 義晴| 杉田 勝彦|
抄録 In this paper we compared the balium preparations having high density and low viscosity (BP-HDLV) with those currently in use. We adopted the stomaches of pigs and the new phantom manufactured by ourselves. We especialy noted the quality of the contrast. The results indicate that BP-HDLV do not always make high contrast. The contrast is related to fluidity besides density. The fluidity of BP-HDLV should be improved. In adition, it costs much to increase content of balium. Therefore, we can not find out positive significance to use these BP-HDLV.
キーワード contrast medium balium density viscosity fluidity
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1994-01-31
4巻
開始ページ 57
終了ページ 60
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313918
JaLCDOI 10.18926/11936
タイトル(別表記) The contrast media containing soft drinks for X-ray examination of the stomach
フルテキストURL 005_111_114.pdf
著者 渋谷 光一| 中桐 義忠| 東 義晴| 後藤 佐知子| 杉田 勝彦| 大倉 保彦| 丹谷 延義| 延原 栄太朗| 三上 泰隆|
抄録 胃検診の受診者を増やし、検査を円滑に行う上で、造影剤である硫酸バリウム懸濁液の飲みにくさが障害の一つとなっている。これを改善する試みとして、懸濁液に市販の粉末清涼飲料を混入し検討したところ72.1%の人から通常のものよりも飲みやすいという回答を得た。今後、臨床的な検討を加えながら、さらに創意工夫していく必要がある。
抄録(別表記) Barium sulfate suspensions are disliked to drink by many people. That is one of the obstacles to increse person who consult doctors and to have medical examinations smoothly. We mixed soft drinks into the suspensions to improve this problem. As a result they were popular among 72.1% people. We need to study about these suspensions further, adding the clinical matters.
キーワード 造影剤 (contrast media) 硫酸バリウム製剤 (barium sulfate suspension) 粘度 (viscosity) 飲みやすさ (easiness to drink)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1995-01-31
5巻
開始ページ 111
終了ページ 114
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313324
JaLCDOI 10.18926/11934
タイトル(別表記) Granularity of Asymmetric Screen-Film System
フルテキストURL 005_105_110.pdf
著者 本田 貢| 吉田 彰| 中村 伸枝| 三村 誠一| 門久 繁文| 三上 泰隆| 東 義晴| 竹内 知行| 杉田 勝彦| 平木 祥夫|
抄録 非対称増感紙フィルム系について、両面乳剤フィルムのウィナースペクトル測定に及ぼす影響を検討した。濃度0.5~2.5のノイズ試料のフロント乳剤、バック乳剤、支持体層それぞれで測定したスペクトルの和と、両面同時に測定した全スペクトルとを比較した。その結果、全ウィナースペクトルの値は、フロント乳剤、バック乳剤の和より高くなった。また、試料濃度が高いほど、その差は大きくなった。これらは、マイクロデンシトメータの光学系の配置と両面乳剤フィルムの構造に起因している。したがって、非対称システムのようにフィルムの前面と後面で特性を分けて考える場合、注意を要する。
抄録(別表記) We have examined a factor affecting the Wiener spectrum measurement of double-emulsion film. An asymmetric screen-film system (Kodak HC/INSIGHT system) was used in this study. The results indicated that the total Wiener spectra of double-emulsion films are higher than the sum of spectra obtained with the front emulsion and back emulsion separately. These differences are attributed to the arrangement of the optical system of the microdensitometer used to measure and the structure of double-emulsion film.
キーワード 非対称増感紙フィルム系 (asymmetric screen-film system) 両面乳剤フィルム (double-emulsion film) ウィナースペクトル (Wiener spectrum) 粒状性 (granularity) マイクロデンシトメータ (microdensitometer)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1995-01-31
5巻
開始ページ 105
終了ページ 110
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313805
JaLCDOI 10.18926/11931
タイトル(別表記) Statistics and analysis on the troubles of linear accelarator.
フルテキストURL 005_099_103.pdf
著者 中桐 義忠| 三上 泰隆| 東 義晴| 渋谷 光一| 後藤 佐知子| 稲村 圭司| 田原 誠司| 宇野 弘文| 杉田 勝彦| 平木 祥夫|
抄録 放射線治療の成否は厳密に設定されたTarget Volumeに如何に正確な線量を照射するかによって決まる。治療術式の過程において、最も大きな誤差を生む要因は照射機器である。誤差の少ない治療を目指す第一歩は機器を正確に作動させることであり、日常からの保守点検および整備が必要である。今回、岡山大学附属病院で1976年から1991年までに使用されたリニアックについて、その故障状況を集計し、部位別故障件数、管球の寿命、稼動率などを分析検討した。その結果、故障件数では設置され稼動を始めた1976年、装置の老朽化が進んだ1990、1991年に多かった。部位別集計では加速部に圧倒的に多く、次いで照射口、高圧部の順であった。稼動率は設置年および1987年を除いてはいずれも96%以上とよい結果であった。この結果は全国に稼動している同型の装置の保守点検に役立つものと考える。
抄録(別表記) The Accuracy of radiation dose exposed the rigidly selected target volume is one of significant factors that have an influence upon the efficiency of treatment in the radiotherapy techniques. Therefore, it is necessary to daily maintain and check irradiation equipment for the radiotherapy. For the radiotherapy, the electron linear accelerator, Toshiba LMR-15A, had worked from 1976 to 1991 in Okayama University Hospital. On this study, all records regarding the operating condition were analyzed concerning parts of trouble, life of a tube and operating efficiency for these sixteen years. In a high frequency of trouble, an accelerating structure ranked first, followed by a collimator and a high voltage generator. The operating efficiency was 95% or greater. This report should be helpful to maintain and check a same model in other hosptals.
キーワード 放射線治療 (radiotherapy) 直線加速器 (linear accelarator) 故障 (troubles of radiation system)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1995-01-31
5巻
開始ページ 99
終了ページ 103
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313722
JaLCDOI 10.18926/11900
タイトル(別表記) Faculties' ideas to the school syllabus
フルテキストURL 005_001_008.pdf
著者 太田 武夫| 下石 靖昭| 東 義晴| 遠藤 浩| 岡﨑 愉加|
抄録 シラバスは、その内容も含めて、まだ教官にも、学生にもなじみのない言葉である。著者らは、1994年、初めてのシラバス(講義要覧)を刊行した。このシラバスの中では、1人もしくは数人の教官によって、1教科・1学期分の講義の概要が各1ページの中に記載され、その枠組みは、基本的事項以外に、抗議の目標、授業計画、授業計画、教科書、参考書、授業の進め方、評価の方法、教官のメッセージから成るものであった。このシラバス作成に参加した専任教官45人及び非常勤教官107人、計153人に対し、調査表によってシラバス作成に関する意見を求めた結果、117人(77.0%)の教官が解答を寄せた。その結果は次のようなものであった。1.今回の作成計画以前にシラバスについて認識があった者は44.5%であった。しかし、シラバスの今後の発行に関しては94.9%の者が賛成であった。今回の枠組みについてはほとんどの者が肯定的であった。2.学生による授業評価については、約半数(52.7%)が賛成であるが、時期尚早とする者が41.8%いた。
抄録(別表記) The term 'syllabus' and its concept are not familiar to university faculty members and students in Japan. Authors who belonged to School of Health Sciences Okayama University, published the first syllabus in 1994. In this syllabus, an outline of each course for a semester was shown by one or a few faculty members. The framework consisted of the mark of the subject, schedule, textbook, reference book, how to teach, how to grade and message to students, besides basic information. Forty-five full-time faculty members and 107 part-time faculty members were asked about their ideas for the syllabus through a questionnaire survey by mail and 119 faculty members (78.3%) responded to it. Only fifty-three faculty members (44.5%) had known the term 'syllabus' before the time when they started to talk about it or were asked to write it. But after their presenting syllabi, 104 faculty members (87.4%) agreeded to continue publishing and they were affirmative to framework as it was proposed this time. As for evaluation of subject by students, about half of faculties (52.7%) agreeded and 51.8% of them replied that were not mature enough for it yet.
キーワード シラバス (syllabus) 授業要覧 授業概要 教官の意見 (faculties' idea) evaluation by student
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1995-01-31
5巻
開始ページ 1
終了ページ 8
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313605
著者 渋谷 光一| 中桐 義忠| 東 義晴| 杉田 勝彦| 小橋 高郎| 大倉 保彦| 丹谷 延義| 三上 泰隆| 平木 祥夫|
発行日 1995-06-30
出版物タイトル 岡山医学会雑誌
107巻
1-6号
資料タイプ 学術雑誌論文
著者 中桐 義忠| 東 義晴| 渋谷 光一| 後藤 佐知子| 丸山 敏則| 杉田 勝彦| 園山 真世| 森岡 泰樹| 稲村 圭司| 田原 誠司| 宇野 弘文| 門久 繁文| 井上 龍也| 平木 祥夫|
発行日 1995-12-20
出版物タイトル 環境制御
17巻
資料タイプ 紀要論文
JaLCDOI 10.18926/11733
タイトル(別表記) Physical Properties of Screen-Film Systems - New System for Chest Radiography -
フルテキストURL 006_007_013.pdf
著者 後藤 佐知子| 吉田 彰| 中村 伸枝| 丸山 敏則| 東 義晴| 渋谷 光一| 中桐 義忠| 杉田 勝彦| 本田 貢|
抄録 胸部専用の新しい増感紙/フィルムシステムについて物理的画質特性を測定し、従来のシステムと比較を行った。測定した新システムは、日本コダックのインサイトシステムからHC/IT-1、富士メディカルシステムのADシステムからHG-M/UR-1、コニカのEXシステムからXG-S/ES-C である。従来のシステムとして富士メディカルシステムのHR-4/Super HR-S を使用した。その結果、新システムは22~73%感度は高く、特性曲線においては低濃度部を持ち上げ、最大コントラストは高濃度側にシフトしていた。空間周波数2cycles/mmで相対鮮鋭度を比較すると、新システムが10~30%程度の低下となっていた。ノイズは、濃度1.0で新システムが35~46%の減少となった。以上より、新システムは、縦隔部の濃度を上げ、粒状性の改善を重視した設計となっていることが分かった。
抄録(別表記) The new three screen-film systems, Kodack INSIGHT system HC/IT-1, Fuji AD system HG-M/UR-1, and Konica EX system XG-S/ES-C for chest radiography were evaluated. In addition, the conventional system Fuji HR-4/Super HR-S was measured for comparison with new systems. The evaluation method is as follows. The image quality characteristics, speed, contrast, resolution and granularity were measured. The sensitivity and contrast were determined by measuring the MTF (modulation transfer function) and NCTF (normalized contrast transfer function). The granularity was estimated by measuring the Wiener spectrum. The speed of the new systems were improved by 22% to 73%. The dinsities at the maximum gradient value were shifted to high density (2.0-2.7). However, they had 10% to 30% less resolution. The other hand, the granularities were improved by 21% to 37% at 1.0 density. These results suggested that the emphasis of new systems was put on improving the image quality of mediastinum region, maintaining performance of the conventional system.
キーワード 増感紙フィルム系 (screen-film system) 物理特性 (physical properties) 感度 (speed (film sensitivity)) コントラスト (contrast) 鮮鋭度 (resolution) 粒状性 (granularity)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1996-02-29
6巻
開始ページ 7
終了ページ 13
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313511
JaLCDOI 10.18926/11735
タイトル(別表記) Clinical application of the contrast media containing soft drinks for the X-ray examination of the stomach
フルテキストURL 006_015_021.pdf
著者 渋谷 光一| 中桐 義忠| 東 義晴| 後藤 佐知子| 丸山 敏則| 杉田 勝彦| 大倉 保彦| 夏川 浩一| 丹谷 延義| 延原 栄太朗| 森岡 泰樹| 竹田 芳弘| 平木 祥夫|
抄録 胃X線検査の被験者を増やし、検査を円滑に行う上で造影剤である硫酸バリウム懸濁液の飲みにくさが障害の一つとなっている。これを改善する目的で、懸濁液に市販の粉末清涼飲料を混入したところ、64.2%の人が通常のものよりも飲みやすいと答えた。臨床上も造影剤の付着が悪くなったり、胃の蠕動を促進する等の問題は生じず、有用であると考えられた。
抄録(別表記) Barium sulfate suspensions are disliked by many people. This is one of the obstacles to increase persons who consult doctors and to have medical examinations smoothly. We mixed soft drinks into the suspensions to improve this problem. As a result, they were popular among 64.2% people, there were not any clinical troubles that the suspensions did not stick to the mucous membrane of the stomach well, the peristalsis of the sotmach were promoted, and so on. These results suggest that the contrast media containing soft drinks is useful for the X-ray examinations of the stomach.
キーワード 造影剤 (contrast media) 硫酸バリウム (barium sulfate suspension) 清涼飲料 (soft drinks) 消化管検査 (barium examination)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1996-02-29
6巻
開始ページ 15
終了ページ 21
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313345
JaLCDOI 10.18926/11766
タイトル(別表記) Basic properies of beam flatness analyzer (THEBES 7000)
フルテキストURL 007_1_023_027.pdf
著者 中桐 義忠| 小栗 宣博| 丸山 敏則| 大川 義弘| 若狭 弘之| 永谷 伊佐雄| 東 義晴| 渋谷 光一| 後藤 佐知子| 三上 泰隆| 稲村 圭司| 森岡 泰樹| 杉田 勝彦| 平木 祥夫|
抄録 癌の有力な治療法の1つである放射線治療はシステム全体の許容誤差が±5%と、正確な治療が要求されている。正確な線量を正確な場所に照射されなければならない。そのためには日頃から照射装置のQuality Control が重要である。Beam平坦度の測定はその中でも、重要な項目で1/6月の測定頻度が勧告されている。今回、市販のBeam Flatness Analyzer (7000型 THEBES) を使用する機会を得たので、その基礎的性能であるChamber 相互の感度のちがい、測定値の再現性、識別可能な最小線量について実験を行った。各項目とも誤差が非常に少なく、日常の使用に際してそのデータは十分信頼できるものであることがわかった。
抄録(別表記) Lrradiation, which is one of the effective therapies to Cancer, is required to be accurate within ±5% of the tolerable error of the total system. An accurate dose has to be irradiated accurately at a target volume. For that Quality Control of the irradiation equipment is cotinually important. As the measurement of Beam Flatness is an important item, it is obligated to be measured once a week. In this paper as we have an opportunity to use a Beam Flatness Analyzer sold on a corporation of medical instrument, we experiment about the difference of sensitivity of mutual Chamber which is its basic property, the reproducibility of the measured score, and the minimum value of measuralbe dose. We find that the difference in every item is so small that its data is very reliable in a daily use.
キーワード 放射線治療 (radiation therapy) ビーム平坦度 (beam flatness) 性能評価 (valuation of property) 電子加速器 (electron accelerator)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1996-09-30
7巻
1号
開始ページ 23
終了ページ 27
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313499
JaLCDOI 10.18926/11770
タイトル(別表記) A study on students' idears about the school syllabus
フルテキストURL 007_1_029_033.pdf
著者 太田 武夫| 下石 靖昭| 東 義晴| 一村 光子| 遠藤 浩| 岡﨑 愉加|
抄録 1995年4月に初めての本学のシラバス(講義要覧)が発行され、学生に配布された。本報は約1年後の1996年2月の試験期間中にシラバスに対する学生の意見を問うために行われたアンケート調査の結果をまとめたものである。回答を寄せた学生は435人で、在籍学生の89.0%であった。よく利用したとする学生は18.2%、数回利用したとする学生は67.6%であった。学生の使用目的はテストの準備や履修科目の決定のためであった。以上のことから、現在のシラバスは学生が日常的により活用するためにはさらに教師が改善、工夫する必要があると思われた。
抄録(別表記) The first syllabus of the School was published and distributed to all of the students in April,1995. Ten months later during the term examination, an opportunity arouse to question the sudents about their idears for the syllubus. A survery of all of the students was carried out and 435 students (89.0%) responded to it. They generally utilized it well, especially to decide which subjects they will take, to find out about the textbooks that will be used, and how to prepare for the examination. However it was found not to be sufficient for their daily use. Teachers therefore must make a greater effort in giving information about their lecture.
キーワード シラバス (syllabus) 授業計画 講義要覧 学生の意見 (student) use
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1996-09-30
7巻
1号
開始ページ 29
終了ページ 33
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002313997
JaLCDOI 10.18926/15282
タイトル(別表記) Influence of the Bone for Electron-beam Therapy
フルテキストURL 008_1_037_042.pdf
著者 中桐 義忠| 稲村 圭司| 田原 誠司| 宇野 弘文| 東 義晴| 渋谷 光一| 後藤 佐知子| 丸山 敏則| 森岡 泰樹| 栄 勝美| 中川 富夫| 武本 充広| 杉田 勝彦| 平木 祥夫|
抄録 電子線治療は,体表面またはその近傍に発生する悪性腫瘍の治療にしばしば用いられる治療法である。電子線はエネルギーに対応した飛程を持っており,飛程を越えると急激に線量は減少する。この性質は腫瘍に一定の線量を照射し,腫瘍後方に存在する決定臓器を保護することができるので,病巣を選択的 に治療するのに好都合である。しかし,照射野内に人体軌部組織より密度の違う物質が存在する場合,散乱,吸収の影響が大きく,電子線線量分布は乱れたものとなる。今回,人体内にある骨を想定してVolumeの違う骨Phantomを使 用して影響を調べたところ,骨幅によって骨後方および断端に線量の乱れが生じることがわかった。すなわち,骨の中央ではある程度後方に距離が離れると,線量は大きく減少する現象が見られた。また,横方向の線量分布は骨断端近くで一旦線量の減少が見られ,断端を離れると急激に増加する。したがって,実際の臨床において,Target Volume近くに骨が存在する場合は総線量の決定に際して注意が必要である。
抄録(別表記) The Electron-beam therapy is often used for the treatment to malignant tumors on the surface or its neighborhood of the body. An electron-beam has the range depended on its energy, and its dose is rapidly decreased at the points beyond the range. This property is available to irradiating a tumor with leaving the critical organs unexposed. However, we should take notice that the dose-distributions are disordered if materials of various densities are in the irradiation field. In this paper, we examined how the dose-distribution are disordered by using the phantom made from human bones of various volumes. As a result, the peculiar disorders of the dose-distribution due to the difference of the bone width were found at the back and edge of a bone. Moreover, the dose at the edge of a bone was decreased, but the dose at the side of a bone was rapidly increased in the distance position from the bone. It is necessary to take account of that the above peculiar phenomenon affect the total dose for the treatment when bones exist near the target volume in the irradiation field.
キーワード 電子線治療 (electron-beam therapy) 線量分布 (dose distribution) 不均質物質 (inhomogeneity)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1997-09-10
8巻
1号
開始ページ 37
終了ページ 42
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002307839
JaLCDOI 10.18926/15261
タイトル(別表記) 現像温度および時間のマンモフイルム特性に与える影響
フルテキストURL 008_2_121_129.pdf
著者 丸山 敏則| 後藤 佐知子| 東 義晴| 中桐 義忠| 渋谷 光一| 杉田 勝彦| 竹内 知行| 本田 貢| 吉田 彰|
抄録 The influence for developing temperature and processing time within film processing conditions was investigated using four mammographic films, Konica New CM, Fuji UM-MA HC, Kodak Min-R M and Kodak EB/RA (for rapid system). And Fuji UR-2, a double-emulsion film, was used as a control. Those sensitometric strips exposed by a sensitometer were processed in the different combinations of developing temperatures ranging from 28 to 36℃, processing times from 45 to 210 sec. Average gradient, relative speed and base plus fog obtained from the measured film characteristic curves were evaluated for the different developing temperatures and times. Fuji UR-2 was scarcely affected and mammographic films were greatly affected in the different combinations without an increase in base plus fog except EB/RA. In New CM, UM-MA HC and Min-R M, the average gradients and the relative speeds increased as the developing temperature was higher and the developing time was longer, but the increases were limit on the combination of 36℃ and 210 sec in New CM and UM-MA HC. In EB/RA, the average gradients were almost constant and the relative speeds increased slightly like the double-emulsion film. These results suggested that it would be possible to contribute to dose reduction and advancement of contrast in New CM, UM-MA HC and Min-R M by changing these processing parameters.
抄録(別表記) フィルム処理条件において,現像温度と処理時間に対する影響を4種類のマンモグラフィ用フィルムKonica New CM, Fuji UM-MA HC,Kodak Min-R M,迅速処理用Kodak EB/RAについて調べた。そして,比較基準用として両面乳剤フィルムFuji UR-2を用いた。感光計で露光したフィルムを現像温度28~36℃,処理時間45~210秒で処理した。特性曲線から得られたフィルム特性(平均階調度,相対感度,カブリ濃度)を異なる現像温度,現像時間に対して評価した。UR-2はほとんど影響を受けず,マンモグラフィ用フィルムは,カブリ濃度が上昇することなく,現像条件の影響を大きく受けた。New CM, UM-MA HC,Min-R Mは現像温度の上昇,処理時間の延長に伴い,平均階調度と相対感度は増加した。しかし,New CM, UM-MA HCの36℃,210秒で増加は限度に達した。EB/RAの平均階調度は一定で,相対感度は両面乳剤フイルムと同 様にわずかな増加であった。これらの結果は,New CM, UM-MA HC, Min-R Mにおいて,処理条件を変化させることにより,被曝低減,コントラスト向上に貢献できる可能性を示唆していた。
キーワード マンモグラフィ (Mammography) 現像処理パラメータ (Processing Parameter) 特性曲線 (Characteristic Curve) 平均階調度 (Average Gradient) 相対感度 (Relative Speed)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1998-01-31
8巻
2号
開始ページ 121
終了ページ 129
ISSN 0917-4494
言語 English
論文のバージョン publisher
NAID 120002307943
JaLCDOI 10.18926/15260
タイトル(別表記) Examination of density and mixing ratio of barium preparations, using an originally created phantom
フルテキストURL 008_2_113_119.pdf
著者 渋谷 光一| 宮原 智則| 鷲見 和幸| 大倉 保彦| 三宅 浩一| 奥田 和裕| 丹谷 延義| 東 義晴| 後藤 佐知子| 丸山 敏則| 中桐 義忠| 杉田 勝彦| 平木 祥夫|
抄録 胃Ⅹ線検査に造影剤として用いられる硫酸バリウム懸濁液の濃度を,簡便に,客観的に決定する目的で,歯科用アルギン酸塩印象材を用いて,コインの図柄を写しとったファントムを制作した。これを使って二種類の硫酸バリウム製剤の適正濃度(PD)を調べた結果,BARITOP PとBARICON MEALを単体,若くは混合した場合には,おおよそ,PD(W/V%)=0.75C+165(ただし,CはBARICON MEALの混合比(%))となった。また,BARICON MEALはコントラストが高く,精密検査に有利だと考えられ,BARITOP PとBARICON MEALを混合すると濃度の許容範囲が広がるため,通常の検査に都合がよいと考えられた。更に,このファントムは簡便に作成することができ,しかも,硫酸バリウム懸濁 液との親和性も良いことから,硫酸バリウム製剤やⅩ線TV装置の評価,増感紙・フィルム系の評価等にも応用できると考えられた。
抄録(別表記) A phantom copied designs of coins was created with alginate dental impression material to establish a objective, simple and easy method for deciding density of barium sulfate suspensions which are used for stomach X-ray examination as contrast media. As a result of examining proper density (PD) of two kind of barium preparations with this phantom, in case of BARITOP P and BARICON MEAL, PD were almost shown by the next equation : PD(W/V%)=0.75C+165(C meant content of BARICON MEAL (%)). As BARICON MEAL resulted in hight contrast, it was thought to be suitable for a close examination. Mixture of BARITOP P and BARICON MEAL was thought to be suited to a routine examination becouse of wide permissible level of density. This phantom would be able to apply to the test of balium preparations, X-ray TV systems, screen/film systems and so on.
キーワード 造影剤 (contrast media) 硫酸バリウム (barium sulfate suspension) 胃X線検査 (X-ray examination of the stomach) アルギン酸塩印象材 (alginate impression material) ファントム (phantom)
出版物タイトル 岡山大学医療技術短期大学部紀要
発行日 1998-01-31
8巻
2号
開始ページ 113
終了ページ 119
ISSN 0917-4494
言語 Japanese
論文のバージョン publisher
NAID 120002307835