タイトル(別表記) | On the measurement of oxygen and carbon isotopic ratios of carbonates
|
フルテキストURL | |
著者 |
渡辺 洵
東京大学理学部地質学教室
松葉谷 治
岡山大学温泉研究所温泉化学部門
|
抄録 | Oxygen and carbon isotopic ratios of two working standards of our laboratory and three other standards were measured by McCrea's technique
and the new McKINNEY type mass spectrometer in
this Institute. The reproducibility of δ(13)C measurement is about ±0.2‰. Some systematic differences in δ(18)O values were observed among the three measurements in Apr.-May, July, and Aug., 1971. Because these systematic differences are supposed to be caused by the imperfection
in the processes of decomposition of carbonates
by phosphoric acid, the reproducibility of δ(18)O
measurement may be ±O.2‰ in most cases. The calibrations of our working standard, CK-13, against the PDB standard were carried out on the assumption that δ(18)O and δ(13)C of CK-13 are -1.75‰ and +O.54‰ relative to PDB, respectively. The accuracy of these calibrations is about ±O.2‰ as shown by the interlaboratory comparison of some standards (Table 3). The δ(18)O values of the two working standards relative to PDB are independently recalculated relative to SMOW based on the two assumptions (Table 4). One of which is that a standard water MSA-2 is -8.20‰ relative to SMOW, and the other is that the δ(18)O of CK-13 is -1.75‰ relative to PDB. The results of both calculations agree within 0.4‰.
|
備考 | 原著論文 (Original Paper)
|
出版物タイトル |
岡山大学温泉研究所報告
|
発行日 | 1972-03-25
|
巻 | 41巻
|
出版者 | 岡山大学温泉研究所
|
出版者(別表記) | Institute for Thermal Spring Research, Okayama University
|
開始ページ | 9
|
終了ページ | 12
|
ISSN | 0369-7142
|
NCID | AN00032853
|
資料タイプ |
紀要論文
|
OAI-PMH Set |
岡山大学
|
言語 |
日本語
|
論文のバージョン | publisher
|
NAID | |
Eprints Journal Name | mmc
|