このエントリーをはてなブックマークに追加


ID 16537
Eprint ID
16537
フルテキストURL
93_315.pdf 554 KB
タイトル(別表記)
A study on the position of chest electrodes for the Frank lead system Part I. A study using a human torso model
著者
宮阪 実 岡山大学医学部第一内科教室
抄録
The effect of horizontal displacement of chest electrodes C, A and I for the Frank lead system on transfer impedance was studied in a human torso model. The results obtained were as follows: 1) With electrode C displaced 2 cm to the right of its original position, the magnitude of transfer impedance of lead Z increased by 6.3% and the azimuthal angle of transfer impedance of lead X deviated by 3.7%. A displacement of 2 cm to the left caused a decrease in the magnitude of the transfer impedance of lead Z of 5.6%. On the other hand, backward deviation changed the azimuthal angle of the transfer impedance of lead Z by 2.8°. 2) By moving electrodes A and I forward simultaneously by 2 cm the magnitude of the transfer impedance of lead X increased by approximately 5% while 2 cm backward-displacement of the electrodes decreased the magnitude of the transfer impedance of lead X by 5.7%. 3) With the center of the heart approaching the precordial wall, changes observed in 1) and 2) became conspicuous when the electrodes were on the left but obscure when the electrodes were located backwards or to the right.
キーワード
Frank誘導法
胸部電極
人体胴体模型
誘導ベクトル
X, Y, Z誘導
発行日
1981-04-30
出版物タイトル
岡山医学会雑誌
出版物タイトル(別表記)
Journal of Okayama Medical Association
93巻
3-4号
出版者
岡山医学会
出版者(別表記)
Okayama Medical Association
開始ページ
315
終了ページ
326
ISSN
0030-1558
NCID
AN00032489
資料タイプ
学術雑誌論文
オフィシャル URL
https://www.jstage.jst.go.jp/article/joma1947/93/3-4/93_3-4_315/_article/-char/ja/
関連URL
http://www.okayama-u.ac.jp/user/oma/index.html
言語
日本語
著作権者
岡山医学会
論文のバージョン
publisher
査読
有り
Eprints Journal Name
joma