抄録

A single sheet tester having closed magnetic path (a closed type of SST) has a problem that measurement accuracy of magnetostriction is considerably affected by electromagnetic force between specimen and yoke. Therefore, an open type has been developed. In order to get uniform flux distribution in a sufficiently large region, a compensating magnetizing winding is installed, and a method of waveform control is investigated, in which applied voltages to main and compensating windings are adjusted individually. The effectiveness of the newly developed open type is demonstrated by measuring magnetostrictions of thin amorphous sheet as well as highly grain-oriented silicon steel sheet