ID | 15789 |
JaLCDOI | |
Sort Key | 7
|
フルテキストURL | |
著者 |
Korenaga Takashi
Center for Environmental Science and Technology, Okayama University
Moriwake Tosio
Department of Synthetic Chemistry
Takahashi Teruo
Department of Industrial Chemistry
|
抄録 | Three methods for determining chemical oxygen demand (COD) by means of flow injection analysis (FIA) with potassium permanganate, potassium dichromate, or cerium(IV) sulfate as oxidant, developed in this laboratory, are described from the point of view of their operating properties. The permanganate method is the most sensitive and common, but forms manganese(IV) oxide precipitate which blocks the FIA lines and connectors. Addition of phosphoric acid in the reagent system is, however, effective to some extent in order to avoid blocking the flow system. The dichromate method is the most workable and stable, but produces toxic wastes containing chromium(VI). The cerium method is the most probable because cerium(IV) is the strongest oxidant of the three and less poisonous. The last method with
cerium(IV) sulfate is therefore recommendable to apply to the continuous monitoring of COD in many situation.
|
出版物タイトル |
Memoirs of the School of Engineering, Okayama University
|
発行日 | 1984-11-26
|
巻 | 19巻
|
号 | 1号
|
出版者 | 岡山大学工学部
|
出版者(別表記) | School of Engineering, Okayama University
|
開始ページ | 53
|
終了ページ | 62
|
ISSN | 0475-0071
|
NCID | AA00733903
|
資料タイプ |
紀要論文
|
OAI-PMH Set |
岡山大学
|
言語 |
英語
|
論文のバージョン | publisher
|
NAID | |
Eprints Journal Name | mfe
|