Studies on 24-hour Holter electrocardiography Part Ⅱ: Pacemaker function, arrhythmias and hemodynamics of cases with permanent pacemaker implantation
著者
畑 隆登 岡山大学医学部第一外科学教室
抄録
One hundred twenty-four patients with unipolar permanent pacemaker implantation (advanced atrioventricular block, 62 cases; sick sinus syndrome, 51 cases; and atrial fibrillation with slow ventricular response, 11 cases) were studied by 24-hour Holter electrocardiography. Analysis of pacemaker function and arrhythmias with various pacing modes (VVI pacing, 70 cases; AAI pacing, 19 cases; and DDD pacing, 35 cases) and measurement of hemodynamics revealed the following. In all pacing modes, atrial contraction was observed in cases of advanced atrioventricular block. However, in the VVI pacing mode with sick sinus syndrome, atrial contraction was often lost. In sick sinus syndrome group Ⅲ, pacemaker non-mediated arrhythmias were often recorded. Pacing failure was not noted with any atrial or ventricular pacing. Undersensing was observed in 26% of the AAI pacing group and 7% of the VVI pacing group. Myopotential inhibition was observed in 11-24% of all pacing modes, but only two patients complained of subjective symptoms. DDD pacing with atrial sensing yielded effective increase of heart rate, but caused pacemaker tachycardia and pacemaker mediated tachycardia. In making these diagnosis and managing of pacemaker mediated arrhythmias, 24-hour Hotter recordings were the most effective. In comparison of hemodynamics in any pacing mode, physiological pacing was significantly different from VVI pacing as to blood pressure, left atrial pressure and cardiac index. Patients who require improvement of hemodynamics should be treated with physiological pacing. As for the optimal choice of pacing modes, patients with advanced atrioventricular block without underlying disease do not always require physiological pacing. However, the patients with sick sinus syndrome should receive physiological pacing.
キーワード
ホルター心電図
ペーシングモード
ペースメーカー作動状態
ペースメーカー合併不整脈
ペースメーカー血行動態
発行日
1986-06-30
出版物タイトル
岡山医学会雑誌
出版物タイトル(別表記)
Journal of Okayama Medical Association
98巻
5-6号
出版者
岡山医学会
出版者(別表記)
Okayama Medical Association
開始ページ
563
終了ページ
576
ISSN
0030-1558
NCID
AN00032489
資料タイプ
学術雑誌論文
オフィシャル URL
https://www.jstage.jst.go.jp/article/joma1947/98/5-6/98_5-6_563/_article/-char/ja/
関連URL
http://www.okayama-u.ac.jp/user/oma/index.html
言語
日本語
著作権者
岡山医学会
論文のバージョン
publisher
査読
有り
Eprints Journal Name
joma