| ID | 32842 |
| JaLCDOI | |
| フルテキストURL | |
| 著者 |
Kato, Yasuhiro
Hiroshima University
Mawatari, Hideo
Hiroshima University
Nishimura, Shin-Ichiro
Hiroshima University
Sakura, Nobuo
Hiroshima University
Ueda, Kazuhiro
Hiroshima University
|
| 抄録 | Several specific assays have been developed for the measurement of etoposide in biological fluids. As large samples are required for high sensitivity, these systems are not appropriate for a pediatric practice. In the present study, however, an improved method for the determination of serum levels of the anticancer drug etoposide was developed, using high-performance liquid chromatography with fixed-wavelength ultraviolet detection. Etoposide was extracted from serum using dichloromethane. The efficiency of extraction from serum was 85.7 +/- 7.7% for etoposide and 81.1 +/- 8.4% for diphenylhydantoin, the internal standard. The serum concentrations of etoposide were measured in 0.2-ml serum samples. The lower limit of detection was 50 ng/ml. Each measurement was completed within 5 min. The linear quantitation range for etoposide was 0.05-50 microg/ml. This assay presents an alternative method for routine measurement of serum levels of etoposide in the pediatric oncology setting. |
| キーワード | etoposide
serum concentration
reversed-phase high-performance liquid chromatography
|
| Amo Type | Article
|
| 出版物タイトル |
Acta Medica Okayama
|
| 発行日 | 2003-02
|
| 巻 | 57巻
|
| 号 | 1号
|
| 出版者 | Okayama University Medical School
|
| 開始ページ | 21
|
| 終了ページ | 24
|
| ISSN | 0386-300X
|
| NCID | AA00508441
|
| 資料タイプ |
学術雑誌論文
|
| 言語 |
英語
|
| 論文のバージョン | publisher
|
| 査読 |
有り
|
| PubMed ID | |
| Web of Science KeyUT |