| ID | 21160 |
| JaLCDOI | |
| Sort Key | 5
|
| タイトル(別表記) | The determination of uric acid using modified Patel's method
|
| フルテキストURL | |
| 著者 |
西村 佳子
岡山大学温泉研究所温泉内科学部門
青木 宏子
岡山大学医学部附属病院三朝分院
御舩 政明
岡山大学医学部附属病院三朝分院
|
| 抄録 | N 13-b method for determination of uric acid by AutoAnalyzer- I was proposed from Technicon Corporation, but the sensitivity of this method was insufficient. According to the modification by Patel, good sensitivity was obtained for determining of uric acid, but insufficient separation between peaks was accompanied. To get good separation, Patel's method was reexamined by the authors, and the results were as follows. Sufficient separation was obtained by modifing the flow diagram of Patel's method. The mean recovery rates of uric acid added to serum and urine were 103.6% and 102.5%, and coefficients of variation were 1.27% and 1.05% respectively. The correlation between this modified Patel's method and the U.A. test Wako method was recognized (serum : n=95, r=0.97, urine : n=35, r =0.99). According to our modified method, uric acid concentration in serum of 68 subjects, were 5.8±0.9mg/100ml in 29 males and 4.4±0.8mg/100ml in 39 females.
|
| 備考 | 原著論文 (Original Papers)
|
| 出版物タイトル |
岡山大学温泉研究所報告
|
| 発行日 | 1980-03-25
|
| 巻 | 50巻
|
| 出版者 | 岡山大学温泉研究所
|
| 出版者(別表記) | Institute for Thermal Spring Research, Okayama University
|
| 開始ページ | 25
|
| 終了ページ | 31
|
| ISSN | 0369-7142
|
| NCID | AN00032853
|
| 資料タイプ |
紀要論文
|
| OAI-PMH Set |
岡山大学
|
| 言語 |
日本語
|
| 論文のバージョン | publisher
|
| NAID | |
| Eprints Journal Name | mmc
|