Fujii, Yasushi Department of Radiology, Chugoku Central Hospital of the Mutual Aid Association of Public School Teachers
Nakayama, Takahiro Department of Radiology, Chugoku Central Hospital of the Mutual Aid Association of Public School Teachers
Oshita, Junki Department of Radiology, Chugoku Central Hospital of the Mutual Aid Association of Public School Teachers
Tsunoda, Ayaka Department of Radiology, Graduate School of Health Sciences, Okayama University
Saeki, Yusuke Department of Radiological Technology, Kawasaki Medical School Hospital
Tanabe, Yoshinori Faculty of Medicine, Graduate School of Health Sciences, Okayama University ORCID Kaken ID researchmap
抄録
Purpose: This study investigates the effect of sagging correction errors on image quality and geometric coordinate accuracy.
Methods: This study utilised the Elekta radiotherapy system, ball bearing (BB), Catphan phantom and MultiMet-WL phantom. Ten distinct flex maps (FMs) were acquired by positioning the BB at the accuracy isocentre and introducing shifts of 0.2, 0.4 and 0.6 mm in the left, table and up directions, respectively. Cone-beam computed tomography images of the Catphan phantom were acquired using 10 FMs. The images were analysed for modulation transfer function (MTF) values and geometric coordinates. Additionally, the Winston–Lutz (W-L) test was conducted under reference couch positions and with a 0.3 mm couch shift.
Results: For the Catphan phantom analysis, the standard deviations of MTF10% across FMs were 0.19. The centre-of-gravity coordinates of the insert exhibited shifts of approximately 0.2, 0.4 and 0.6 mm when comparing reference images to those acquired with the shifted FMs. The results of the W-L test with a 0.3 mm couch shift showed radiation isocentre deviations exceeding 1 mm compared to the reference couch positions.
Conclusions: Minor sagging correction calibration errors did not remarkably impact image quality; however, they altered the geometric coordinates of the image isocentre. These calibration errors decreased the accuracy of off-isocentre positioning.
キーワード
flex map
sagging
Winston–Lutz test
発行日
2026
出版物タイトル
Journal of Radiotherapy in Practice
25巻
出版者
Cambridge University Press (CUP)
開始ページ
e5
ISSN
1460-3969
NCID
AA11974173
資料タイプ
学術雑誌論文
言語
英語
OAI-PMH Set
岡山大学
著作権者
© The Author(s), 2026.
論文のバージョン
publisher
DOI
関連URL
isVersionOf https://doi.org/10.1017/s1460396926100387
ライセンス
https://creativecommons.org/licenses/by/4.0/
Citation
Fujii Y, Nakayama T, Oshita J, Tsunoda A, Saeki Y, Tanabe Y. Effects of sagging correction calibration error on radiation therapy equipment using image analysis. Journal of Radiotherapy in Practice. 2026;25:e5. doi:10.1017/S1460396926100387
助成情報
23K07063: 位相幾何学特徴値を用いた堅固な前立腺強度変調放射線治療法の開発 ( 独立行政法人日本学術振興会 / Japan Society for the Promotion of Science )