このエントリーをはてなブックマークに追加
ID 70184
フルテキストURL
著者
Hirata, Yuichi Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Hiramatsu, Masafumi Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Sugiu, Kenji Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences ORCID Kaken ID publons researchmap
Baba, Fukiko Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Fujita, Juntaro Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Sotome, Yuta Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Kawakami, Masato Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Kimura, Ryu Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Ebisudani, Yuki Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Haruma, Jun Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences ORCID Kaken ID researchmap
Hishikawa, Tomohito Department of Neurosurgery, Kawasaki Medical School Kaken ID publons researchmap
Tanaka, Shota Department of Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
抄録
Background: In recent years, the effectiveness of flow diverters (FDs) for the treatment of intracranial aneurysms has been reported. While FDs are effective, their deployment involves advancing a delivery wire distally, which may pose a risk if a distal aneurysm exists within the same artery. In such cases, the delivery wire could potentially perforate the distal aneurysm. Here, we present two cases of tandem aneurysms in which an internal carotid artery (ICA) aneurysm was treated with an FD following the treatment of a distal cerebral aneurysm.
Case description: A 44-year-old woman and a 67-year-old woman underwent magnetic resonance imaging for headache or abducens nerve palsy. In both cases, two aneurysms were revealed: one at the ICA and the other either at the middle cerebral artery or the top of the ICA. Due to the risk of perforation by the delivery wire during FD deployment, the distal aneurysms were treated first—either with surgical neck clipping or stent-assisted coil embolization. One month after the initial treatment, FD placement for the ICA aneurysm was performed as planned without complications in either case.
Discussion: This is the first report where tandem aneurysms were successfully treated with treatment for distal cerebral aneurysms, followed by FDs for proximal ICA aneurysms. We emphasize the potential risk of perforation of the distal aneurysm by the delivery wire during FD placement.
Conclusion: Treatment of distal cerebral aneurysms beforehand can help ensure the safe and effective use of FDs in patients with tandem aneurysms.
キーワード
Flow diverter
Tandem aneurysms
Complication
Perforation
Delivery wire
発行日
2025-10
出版物タイトル
World Neurosurgery: X
28巻
出版者
Elsevier BV
開始ページ
100540
ISSN
2590-1397
資料タイプ
学術雑誌論文
言語
英語
OAI-PMH Set
岡山大学
著作権者
© 2025 The Authors.
論文のバージョン
publisher
DOI
Web of Science KeyUT
関連URL
isVersionOf https://doi.org/10.1016/j.wnsx.2025.100540
ライセンス
http://creativecommons.org/licenses/by-nc-nd/4.0/