抄録
Correct screw placement is especially difficult in the upper thoracic vertebrae. At the cervicothoracic junction (C7-T2), problems can arise because of the narrowness of the pedicle and the difficulty of using a lateral image intensifier there. Other upper thoracic vertebrae (T3-6) pose a problem for screw insertion also because of the narrower pedicle. We inserted 154 pedicle screws into 78 vertebrae (C7 to T6) in 38 patients. Screws were placed using intraoperative data acquisition by an isocentric C-arm fluoroscope (Siremobile Iso-C3D) and computer navigation. Out of 90 pedicle screws inserted into 45 vertebrae between C7 and T2, 87 of the 90 (96.7%) screws were classified as grade 1 (no perforation). Of 64 pedicle screws inserted into 33 vertebrae between T3 and T6, 61 of 64 (95.3%) screws were classified as grade 1. In this study, we reduced pedicle screw misplacement at the level of the C7 and upper thoracic (T1-6) vertebrae using the three-dimensional fluoroscopy navigation system.
キーワード
Iso-C3D
three-dimensional fluoroscopy
navigation
upper thoracic
pedicle screw
Amo Type
Original Article
出版物タイトル
Acta Medica Okayama
発行日
2010-06
64巻
3号
出版者
Okayama University Medical School
開始ページ
209
終了ページ
212
ISSN
0386-300X
NCID
AA00508441
資料タイプ
学術雑誌論文
言語
英語
著作権者
Okayama University Medical School
論文のバージョン
publisher
査読
有り
PubMed ID
Web of Science KeyUT