ID | 33003 |
フルテキストURL | |
著者 |
Yamamoto, Yoshitake
Okayama University
Adli, Mr.
Okayama University
|
抄録 | Impedance techniques have been widely applied to the biomedical engineering field. In order to obtain definite results from bioimpedance, multi-channel measurement is effective. A linearity of biological tissue and fundamental technical parameters for the instrument must be confirmed. In this study, the fundamental conditions have been investigated for multi-channel bioimpedance measurement and the condition of measurement for biological tissue has been confirmed. The differences of every measurement frequency should be appropriate to 1 kHz and it has been shown that the linearity of biological tissue is maintained. Based on these results, a two-channel bioimpedance measurement instrument has been constructed |
キーワード | bioelectric phenomena
biomedical measurement
electric impedance measurement
|
備考 | Digital Object Identifier: 10.1109/IMTC.1998.679763
Published with permission from the copyright holder. This is the institute's copy, as published in Instrumentation and Measurement Technology Conference, 1998. IMTC/98. Conference Proceedings. IEEE, 18-21 May 1998, Volume 1, Pages 231-234. Publisher URL:http://dx.doi.org/10.1109/IMTC.1998.679763 Copyright © 1998 IEEE. All rights reserved. |
発行日 | 1998-5
|
出版物タイトル |
Instrumentation and Measurement Technology Conference
|
巻 | 1巻
|
開始ページ | 231
|
終了ページ | 234
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
査読 |
有り
|
DOI | |
Submission Path | biology_general/20
|