High Accuracy of Pelvic Tilt Measurement Using Three-Dimensional Planning Software: A Validation Study Using Dry Bone Models

Komuta, Yuki Division of Orthopedic Surgery, Graduate School of Medical and Dental Sciences, Niigata University
Imai, Norio Division of Comprehensive Musculoskeletal Medicine, Graduate School of Medical and Dental Sciences, Niigata University
Homma, Daisuke Division of Orthopedic Surgery, Graduate School of Medical and Dental Sciences, Niigata University
Funayama, Kazuhisa Division of Legal Medicine, Graduate School of Medical and Dental Sciences, Niigata University
Hirano, Yuki Division of Orthopedic Surgery, Graduate School of Medical and Dental Sciences, Niigata University
Horigome, Yoji Division of Comprehensive Musculoskeletal Medicine, Graduate School of Medical and Dental Sciences, Niigata University
Minato, Izumi Department of Orthopedic Surgery, Niigata Rinko Hospital
Kawashima, Hiroyuki Division of Orthopedic Surgery, Graduate School of Medical and Dental Sciences, Niigata University
Publication Date
2026-10
Abstract
The “CR Alignment” function was recently introduced in ZedHip®, the hip joint module of ZedView®, to evaluate pelvic tilt in non-supine positions via two-dimensional (2D)‒3D matching using radiographs. However, its ccuracy has not been validated. This study therefore evaluated the accuracy of pelvic tilt measurements using the CR Alignment function in dry pelvic bone specimens. Two cadaveric pelvic bone models were placed in nine orientations (anterior/posterior tilt and left/right rotation). Both radiographic and CT images were acquired in each position. In Study 1, pelvic tilt measured directly from CT images was defined as the “true value.” Three observers used the CR Alignment function to calculate pelvic tilt on radiographs, and measurement errors were determined relative to the true values. In Study 2, measurement errors were compared between matched (same position) and mismatched (different position) imaging. In Study 1, the mean measurement error was 1.28±0.64°. Intra- and interobserver intraclass correlation coefficients were 0.996-0.998 and 0.997, respectively. In Study 2, the mean errors were 1.20±0.88° (matched) and 1.32±1.03° (mismatched). This software exhibited high accuracy and reproducibility in 2D‒3D matching of pelvic tilt, with lower variability under matched acquisition, demonstrating its utility for pre- and postoperative assessment in joint arthroplasty.
Document Type
Original Article
Keywords
total hip arthroplasty
pelvic tilt
three-dimensional planning software
two-dimensional and threedimensional matching
radiographic analysis