| ID | 54827 |
| FullText URL | |
| Author |
Yoshida, Yukihiro
Department of Orthopedic Surgery, Nihon University School of Medicine
Okamura, Yuki
Department of Orthopedic Surgery, Nihon University School of Medicine
Akita, Mamoru
Department of Orthopedic Surgery, Nihon University School of Medicine
Taniguchi, Masashi
Department of Orthopedic Surgery, Nihon University School of Medicine
Kojima, Toshio
Department of Orthopedic Surgery, Nihon University School of Medicine
Osaka, Eiji
Department of Orthopedic Surgery, Nihon University School of Medicine
Tokuhashi, Yasuaki
Department of Orthopedic Surgery, Nihon University School of Medicine
|
| Abstract | We performed revision surgery in 2 patients for stem fracture of a cemented tumor prosthesis that occurred more than 25 years after the initial surgery. For revision, the global modular replacement system (GMRS) was used. However, as bone cement in the bone could not be adequately removed, stems with respective diameters of 11 and 12.5 mm were used. In revision surgery for cemented tumor prostheses, adequate removal of residual bone cement is optimal. However, when there is a risk of fracture, it may be appropriate to insert a thicker stem after reaming the femoral canal as much as possible, and then fix the stem using the cement-in-cement method.
|
| Keywords | revision
custom-made tumor prosthesis
malignant bone tumor
|
| Published Date | 2017-02
|
| Publication Title |
Acta Medica Okayama
|
| Volume | volume71
|
| Issue | issue1
|
| Publisher | Okayama University Medical School
|
| Start Page | 69
|
| End Page | 72
|
| ISSN | 0386-300X
|
| NCID | AA00508441
|
| Content Type |
Journal Article
|
| language |
English
|
| Copyright Holders | CopyrightⒸ 2017 by Okayama University Medical School
|
| File Version | publisher
|
| Refereed |
True
|
| PubMed ID | |
| Amo Type | Case Report
|
| Submission Path | amo/vol71/iss1/9
|
| JaLCDOI |