このエントリーをはてなブックマークに追加


ID 11808
JaLCDOI
Sort Key
10
Title Alternative
RESOLUTION PROPERTIES OF CRT MONITORS
FullText URL
Author
Takeuchi, Tomoyuki
Yoshida, Akira
Sugita, Katsuhiko
Nakamura, Nobue
Kishi, Kanji
Hiraki, Yoshio
Abstract
We have developed a new technique for measuring the modulation transfer function (MTF) of a cathode ray tube (CRT) monitor. The "nominal" MTF was measured by displaying a line image of a width of one pixel on the CRT monitor using the pattern generator. We also measured the effective pixel width. In order to obtain the "actual" MTF, the nominal MTF was divided by the sinc function which is the Fourier transform of the effective pixel width. The results indicate that the corrected MTF converges to the constant value as the effective pixel width is narrower. Thus, the actual MTFs were determined and compared in the horizontal direction for two CRT monitors.
Keywords
CRT Monitor
Resolution
Effective Pixel Width
Modulation Transfer Function (MTF)
Publication Title
岡山大学医療技術短期大学部紀要
Published Date
1992-02-05
Volume
volume2
Publisher
岡山大学医療技術短期大学部
Publisher Alternative
School of Health Sciences Okayama University
Start Page
81
End Page
88
ISSN
0917-4494
NCID
AN10355371
Content Type
Departmental Bulletin Paper
OAI-PMH Set
岡山大学
language
Japanese
File Version
publisher
NAID
Eprints Journal Name
fhs