| ID | 21085 |
| JaLCDOI | |
| Sort Key | 9
|
| Title Alternative | Automatic measurement of oxygen isotope ratio of water samples
|
| FullText URL | |
| Author |
Chiba, Hitoshi
Kaken ID
researchmap
Sakai, Hitoshi
Yasutake, Masatoshi
|
| Abstract | The automatic sample preparation system for oxygen isotope analysis of natural water samples was constructed. The system is essentially a modification of that originally designed by W. Dansgaard in the University of Copenhagen. Sixty water samples of 5 ml each are automatically equilibrated with CO(2) of 30 ml NTP each within 4.5 hours. The equilibrated CO(2) gases are successively measured for their oxygen isotopic ratios by an automatic mas-sspectrometer, VG903. The time required for the
measurement of the 60 CO(2) samples is about 15 hours; thus approximately 20 hours are required to complete the isotopic measurements of 60 water samples. The accuracy of the oxygen isotopic analyses is about 0.1%. The result of oxygen isotopic analyses by the automatic preparation system agrees with that of the conventional method within about +0.1%. The results indicate that the automatic preparation system is applicable for the oxygen isotope analysis of natural water samples.
|
| Note | 原著論文 (Original Papers)
|
| Publication Title |
岡山大学温泉研究所報告
|
| Published Date | 1985-03-30
|
| Volume | volume56
|
| Publisher | 岡山大学温泉研究所
|
| Publisher Alternative | Institute for Thermal Spring Research, Okayama University
|
| Start Page | 27
|
| End Page | 34
|
| ISSN | 0369-7142
|
| NCID | AN00032853
|
| Content Type |
Departmental Bulletin Paper
|
| OAI-PMH Set |
岡山大学
|
| language |
Japanese
|
| File Version | publisher
|
| NAID | |
| Eprints Journal Name | mmc
|