このエントリーをはてなブックマークに追加


ID 31517
JaLCDOI
FullText URL
fulltext.pdf 1.05 MB
Author
Hayashi, Nobuyoshi
Hayata, Koshi
Sekiba, Kaoru
Abstract

A high performance liquid chromatographic (HPLC) method with electrochemical detection (ECD) was developed for the simultaneous measurement of estrone, estradiol, estriol and estetrol in serum. These hormones were extracted with diethylether, chromatographed on an silica-octadecyl silane (ODS) column with an eluent of phosphate buffer solution-acetonitrile-methanol (volume ratio 152:85:40), and detected by ECD at +1.0V vs. Ag/AgCl. In comparisons between the values measured by this method and radioimmunoassay, significant correlations were noted for estrone (r = 0.759, p less than 0.01), estradiol (r = 0.816, p less than 0.001) and estriol (r = 0.830, p less than 0.001). In clinical applications of this method, differences between cases of the normal and the anencephalic pregnancy in the thirty-eighth week of gestation were distinct not only in the individual estrogen, but also in the profile analysis of estrogens. With this method, all 4 serum estrogens above approximately 500 pg/ml could be measured within 2 h, and the method seemed to be clinically applicable.

Keywords
high perforrmance liquid chromatography
electrochemical detection
serum estrogens
Amo Type
Article
Publication Title
Acta Medica Okayama
Published Date
1985-04
Volume
volume39
Issue
issue2
Publisher
Okayama University Medical School
Start Page
143
End Page
153
ISSN
0386-300X
NCID
AA00508441
Content Type
Journal Article
language
English
File Version
publisher
Refereed
True
PubMed ID
Web of Science KeyUT