| ID | 40148 |
| JaLCDOI | |
| Sort Key | 3
|
| Title Alternative | The effect of pH of water on the measurement of oxygen isotopic ratio by means of CO(2)-H(2)O isotopic exchange technique
|
| FullText URL | |
| Author |
Matsubaya, Osamu
|
| Abstract | The effect of pH of water on the measurement of oxygen isotopic ratio by means of CO(2)-H(2)O isotopic exchange technique was studied. In this study, a new type of reaction vessel was used (Fig. 1). The reaction time for equilibration using this new reaction vessel was nearly equal to that in the previous study (Fig. 2). The oxygen isotopic ratios of each two samples of pH 2, 4, 7, 8, 9, and 10 were measured. There is no significant difference between the observed δ(18)O values of these samples (Table 1). Therefore it is concluded that the effect of pH of water needs not to be taken into account on the measurement of oxygen isotopic ratio by means of CO(2)-H(2)O isotopic exchange technique.
|
| Note | 原著論文 (Original Paper)
|
| Publication Title |
岡山大学温泉研究所報告
|
| Published Date | 1972-03-25
|
| Volume | volume41
|
| Publisher | 岡山大学温泉研究所
|
| Publisher Alternative | Institute for Thermal Spring Research, Okayama University
|
| Start Page | 1
|
| End Page | 2
|
| ISSN | 0369-7142
|
| NCID | AN00032853
|
| Content Type |
Departmental Bulletin Paper
|
| OAI-PMH Set |
岡山大学
|
| language |
Japanese
|
| File Version | publisher
|
| NAID | |
| Eprints Journal Name | mmc
|