Abstract
This study aims to evaluate the output factors (OPF) of different radiation therapy planning systems (TPSs) using a plastic scintillator detector (PSD). The validation results for determining a practical field size for clinical use were verified. The implemented validation system was an Exradin W2 PSD. The focus was to validate the OPFs of the small irradiation fields of two modeled radiation TPSs using RayStation version 10.0.1 and Monaco version 5.51.10. The linear accelerator used for irradiation was a TrueBeam with three energies: 4, 6, and 10 MV. RayStation calculations showed that when the irradiation field size was reduced from 10 × 10 to 0.5 × 0.5 cm2, the results were within 2.0% of the measured values for all energies. Similarly, the values calculated using Monaco were within approximately 2.0% of the measured values for irradiation field sizes between 10 × 10 and 1.5 × 1.5 cm2 for all beam energies of interest. Thus, PSDs are effective validation tools for OPF calculations in TPS. A TPS modeled with the same source data has different minimum irradiation field sizes that can be calculated. These findings could aid in verification of equipment accuracy for treatment planning requiring highly accurate dose calculations and for third-party evaluation of OPF calculations for TPS.
Keywords
Plastic scintillator
Radiation therapy
Small irradiation field
Output factor
Note
This version of the article has been accepted for publication, after peer review (when applicable) and is subject to Springer Nature’s AM terms of use, but is not the Version of Record and does not reflect post-acceptance improvements, or any corrections. The Version of Record is available online at: http://dx.doi.org/10.1007/s13246-024-01438-5
This fulltext file will be available in May 2025.
Published Date
2024-05-16
Publication Title
Physical and Engineering Sciences in Medicine
Volume
volume47
Issue
issue3
Publisher
Springer Science and Business Media LLC
Start Page
1177
End Page
1189
ISSN
2662-4729
Content Type
Journal Article
language
English
OAI-PMH Set
岡山大学
Copyright Holders
© Australasian College of Physical Scientists and Engineers in Medicine 2024
File Version
author
PubMed ID
DOI
Web of Science KeyUT
Related Url
isVersionOf https://doi.org/10.1007/s13246-024-01438-5
Citation
Ando, Y., Okada, M., Matsumoto, N. et al. Evaluation of output factors of different radiotherapy planning systems using Exradin W2 plastic scintillator detector. Phys Eng Sci Med 47, 1177–1189 (2024). https://doi.org/10.1007/s13246-024-01438-5