このエントリーをはてなブックマークに追加


ID 21163
JaLCDOI
Sort Key
8
Title Alternative
Determination of lithium in serum by emission spectrophotometry using "Spectraspan"
FullText URL
Author
Mifune, Masaaki
Abstract
Lithium in serum was determined by a plasma emissionspectrophotometer "Spectraspan" (low power d.c. plasma are, operating on argon). The opitimum condition for ashing of serum by low temperature was studied. The influences of hydrochloric acid and several metallic ions (Na, K, Mg, Ca) on emission intensity were examined. There was a linear relationship between emission intensity and lithium content from 0 to 10μg/ml. Lithum in serum was easily determined by standard addition method. Precision, coefficient of variance and recovery of known amount of lithium added to the sample for 12 replicate analyses were 0.093μg/ml, 1.85%, 99-104%, respectively. Significantly positive correlationship was observed between emission spectrophotometry using spectraspan and atomic absoption spectrophotometry.
Note
原著論文 (Original Papers)
Publication Title
岡山大学温泉研究所報告
Published Date
1980-03-25
Volume
volume50
Publisher
岡山大学温泉研究所
Publisher Alternative
Institute for Thermal Spring Research, Okayama University
Start Page
43
End Page
48
ISSN
0369-7142
NCID
AN00032853
Content Type
Departmental Bulletin Paper
OAI-PMH Set
岡山大学
language
Japanese
File Version
publisher
NAID
Eprints Journal Name
mmc