| ID | 57857 |
| FullText URL | |
| Author |
Watanabe, Atsuyuki
Department of Cardiovascular Medicine, Okayama University
Kaken ID
Fujimura, Osamu
Department of Cardiovascular Medicine, Okayama University
|
| Abstract | This case describes repeated misclassifications of SVT due to AV node reentry as VT by an ICD. This case illustrates the limitations of SVT-VT discrimination algorithm. Careful analysis of the stored tracings is of critical importance to reach the correct diagnosis.
|
| Keywords | Tachycardiav
Atrioventricular node reentry
Cardiac defibrillatorIschemic cardiomyopathy
|
| Published Date | 2020-02-29
|
| Publication Title |
Indian Pacing and Electrophysiology Journal
|
| Volume | volume20
|
| Issue | issue1
|
| Publisher | Elsevier
|
| Start Page | 35
|
| End Page | 38
|
| ISSN | 09726292
|
| Content Type |
Journal Article
|
| language |
English
|
| OAI-PMH Set |
岡山大学
|
| Copyright Holders | © 2019 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V.
|
| File Version | publisher
|
| DOI | |
| Related Url | isVersionOf https://doi.org/10.1016/j.ipej.2019.12.004
|
| License | http://creativecommons.org/licenses/by-nc-nd/4.0/
|
| Citation | Atsuyuki Watanabe, Osamu Fujimura, Repeated misclassifications of tachycardia by an implantable cardiac defibrillator, Indian Pacing and Electrophysiology Journal, Volume 20, Issue 1, 2020, Pages 35-38, ISSN 0972-6292, https://doi.org/10.1016/j.ipej.2019.12.004.
|
| Open Access (Publisher) |
OA
|
| Open Archive (publisher) |
Non-OpenArchive
|