ID | 52655 |
JaLCDOI | |
FullText URL | |
Author |
Akavipat, Phuping
Hungsawanich, Nate
Jansin, Rungruedee
|
Abstract | In neurosurgery in particular, the recommended placement of electrodes for monitoring depth of anesthesia during surgery sometimes conflicts with the surgical site or patient positioning. Therefore, we proposed this study to evaluate the agreement and correlation of bispectral index values recorded from the usual frontal area and the alternate, post-auricular areas in neurosurgery patients. Thirty-four patients scheduled for neurosurgery under general anesthesia were included. Bispectral index (BIS) sensors were placed at both the frontal and post-auricular areas. The anesthesia given was clinically adjusted according to the frontal (standard) BIS reading. The BIS values and impedance were recorded;Pearsonʼs correlation and Bland-Altman plots were analyzed. The bias ± 2SD for the electrode placement before, during, and post-anesthesia were 0±23.32, 1.5±10.69, and 2.1±13.52, while the limits of agreement were -23.3 to 23.3, -12.2 to 9.2, and -17.7 to 13.5, respectively. The correlation coefficient between frontal- and post-auricular-area electrodes was 0.74 with a p-value
<0.001.The post-auricular placement of a BIS electrode is a practical alternative to frontal lobe placement. Nevertheless, proper electrode location is important to minimize error.
|
Keywords | Bland and Altman analysis
electroencephalogram
instrumentation
equipment
|
Amo Type | Original Article
|
Publication Title |
Acta Medica Okayama
|
Published Date | 2014-06
|
Volume | volume68
|
Issue | issue3
|
Publisher | Okayama University Medical School
|
Start Page | 151
|
End Page | 155
|
ISSN | 0386-300X
|
NCID | AA00508441
|
Content Type |
Journal Article
|
language |
English
|
Copyright Holders | CopyrightⒸ 2014 by Okayama University Medical School
|
File Version | publisher
|
Refereed |
True
|
PubMed ID | |
Web of Science KeyUT |