Development of a microcontroller-based dental implant movement checker

クスマ ウイジャヤ サストラ 岡山大学大学院自然科学研究科
岡 久雄 岡山大学医学部保健学科検査技術科学専攻
更谷 啓治 大阪歯科大学有歯補綴咬合学講座
松谷 善雄 大阪歯科大学有歯補綴咬合学講座
龍田 光弘 大阪歯科大学有歯補綴咬合学講座
川添 堯彬 大阪歯科大学有歯補綴咬合学講座
小嶋 英幹 隆祥産業株式会社生産本部技術部
抄録
The aim of this study was to develop a portable device (IM checker) for measuring dental implant mobility using a microcontroller. A constant amplitude and frequency vibration was applied to a dental implant model and the acceleration signal was detected using measuring probe. Then the IM score was obtained using the criteria developed in this study. We made several implant models of different implant lengths, diameters and material of Rigolac® or Molteno®. There was a linear relationship between mechanical mobility at 400 Hz of the models and the IM scores (R(2) = 0.92). The IM checker could discriminate the mobility of dental implant models in twelve measurements with P < 0.01. There was no significant difference in the means of the IM score measured by four operators with P < 0.01. The results indicated that the IM checker had sufficient reliability and could be available in dental clinics.
キーワード
Tooth mobility (歯の動揺)
mechanical mobility (機械モビリティ)
dental implant (インプラント)
manual examination (触診)